Range Switch - HP 400D Operating And Service Manual

Vacuum tube voltmeter
Table of Contents

Advertisement

Section V
Paragraphs 5-9 to 5-16
c.
Rotate mechanicalzero-adjustment screw clock-
wise until meter pointer is tothe left of zero and mov-
ing upscale toward zero.
d.
Continue to rotate
adjustment screw
clockwise;
stop when pointer is exactly on zero.
Ifpointer over-
shoots zero,
repeat steps c and d.
e.
When pointer is exactly on zero,
rotate
adjust-
ment screw approximately
15 degrees
counterclock-
wise.
This
is enough
to free the
zero
adjustment
screw from the meter
suspension.
If pointer
moves
during this step,
because
the
adjustment
screw
is
turned toofar counterclockwise, repeat the procedure
of steps c through e.
5-9.
CABINET
REMOVAL.
a.
Remove the two cabinet retaining screws
at the
rear of the instrument.
b.
Push the instrument chassis forward out of the
cabinet.
The bezel ring remains attached to the front
panel.
c.
When replacing cabinet, pull power cable through
opening at rear of cabinet.
Be sure power cable is
not caught between chassis and cabinet.
Replace re-
taining screws.
5-10.
TUBE
REPLACEMENT.
Do not remove tubes from the voltmeter when
power is applied.
To do so may damage the
voltmeter.
5-11. In many cases instrument
malfunction can be
corrected by replacing
a weak or defectivetube.
Check
tubes
by substitution while
following
the
voltmeter
T.O. 33A1-12-349-1
performance
check procedure
in paragraph 5-22,
Re-
sults obtained through the use of a "tube checker" can
be misleading.
Before
removing the tubes
from the
instrument,
mark
the original
tubes
so they can be
returned to the same
socket if they are not defective.
Replace only those tubes proven to be defective.
5-12. Figure 5-2 lists each tube in the voltmeter with
its function
and the check
or adjustment
required
if
the tube is replaced.
5-13.
REPLACEMENT
OF SPECIAL
PARTS.
5-14.
PRECISION RESISTORS
AND INDUCTORS,
Sev-
eralparts usedinthe voltmeter have closer tolerances
than those
used
in most
test equipment.
Resistors
R104,
R105,
R108,
and R111
through
R116 are pre-
cisioncomponents.
If these resistors require replace-
ment, use the same value and type as the original, as
shown
in the parts
breakdown.
If different
values
are used or component positions are moved, the cali-
bration of the voltmeter may be inaccurate or the fre-
quency response may be altered.
The
inductance
of
L10 and
Lil
affects
the frequency
response
of the
voltmeter.
Do notalter the shape or position of these
coils.
Install
replacement
components
in the same
positions the original
components occupied, as nearly
as possible.
5-15. DIODE RECTIFIERS.
Special high-performance
silicon diodes selected by the Hewlett-Packard
Co.
are used for CR1 and CR2.
When
replacing the sili-
con diodes,
be careful in soldering; heat can damage
them,
Place a heat sink (such as a long-nose pliers)
on each
diode lead close to the diode body to conduct
the heat away.
If CRi and
CR2 are replaced,
the
voltmeter
calibration and frequency response
must
be checked as described in paragraph 5-22.
5-16. RANGE SWITCH.
Because of the critical con-
struction and wiring
of switch
S1, it is not practical
to attempt a major repair onthe switch.
When mech-
anical failure occurs in switchS1,
replace the complete
CIRCUIT
CHECK OR
Cathode
Follower
1st Amplifier
2nd Amplifier
3rd Amplifier
4th Amplifier
High Voltage Rectifier
Series Regulator
Control Tube
Reference Tube
Calibration and frequency response (para, 5-22)
Test of the power supply (para.
5-20)
* Note that Vi must be replaced by a 6CB6,
aged and selected for low noise and microphonics
(4,
Part No. 5080-0621).
Figure 5-2.
Adjustments Required When Tubes Are Replaced
5-2
00102-3

Advertisement

Table of Contents
loading

This manual is also suitable for:

400h400lH02-400d

Table of Contents