Utilities > Dukane > Calibration XDUC
Touch the Calibration XDUC (Position Transducer)
button at the top of the Utilities screen, to activate the
calibration screen shown in Figure 6-39 below.
Figure 6-39 Utilities Screen > Dukane >Calibration XDUC
Calibration Sequence
1. Remove the tool before calibrating.
Use Table Up/Down touch keys to control the table.
2. Raise the table until it stops.
3.
Press the accept button and the right cycle start switch
shown as item H in Figure 4-2.
4. Lower the table until it stops.
5. Press the accept button and the right cycle start switch
shown as item H in Figure 4-2.
Dukane Manual Part No. 403-592-00
Table Up
Voltage Delay –
detect normal operating voltage before the Voltage Over-
load alarm is triggered
Current Delay –
PLC to detect normal operating current before Current
Overload alarm is triggered.
Table Down
Feedback Delay –
PLC to detect a feedback signal, before No Feedback
Detected alarm is triggered.
Input
Section 6 - Touch Screen Menus
NOTE
The position transducer was calibrated at
the factory and normally will not need to be
calibrated again. However, calibration would
be needed for any replacement transducer.
Allow 1 to 3 seconds for the PLC to
.
Allows 1 to 3 seconds for the
Allows 1 to 3 seconds for the
Raw –
Transducer output value being
read from the analog input card.
Volts –
Transducer output voltage.