Overload Measurement Specifications

Typical 'On' Current
Source Voltage
Indication
5 to 9 V
100 to 180 mA

Oscilloscope Connections

With the cable supplied with the SC600 Option, connect the SCOPE output on the
Calibrator to one of the channel connectors on your oscilloscope (see Figure 9-1).
To use the external trigger, connect the TRIG output on the Calibrator to the external
trigger connection on your oscilloscope. To use the external trigger and see its signal with
the calibration signal, connect the TRIG output to A different channel. See your
oscilloscope manual for instructions to connect and see an external trigger.
Figure 9-1. Oscilloscope Connection: Channel and External Trigger

Start the SC600 Option

Push  (LED on) to select the SC600 Option. The SCOPE menu, shown below,
shows in the Control Display. You can push the first four softkeys to go directly to the
VOLT, EDGE, LEVSINE, and MARKER calibration menus. Push the last softkey to go
to the OTHER menu (also shown below), this lets you access to WAVEGEN, VIDEO,
PULSE, Impedance/Capacitance measurement (MEAS Z), and Overload (OVERLD)
menus. Push  to return to the SCOPE menu from the OTHER menu. This chapter
tells you about each of these menus.
SC600 Oscilloscope Calibration Option
Typical 'Off' Current Indication
10 mA
Oscilloscope Connections
Maximum Time Limit DC or AC
(1 kHz)
Setable 1 s to 60 s
5502A
CALIBRATOR
gvx036.eps
9
9-7