Volt Dc Decade Test; Dc Volts Decade Test Configuration - Fluke 515A Technical Manual

Portable calibrator
Table of Contents

Advertisement

TI mV Dial
Setting
1000
700
500
200
100
0
*Connect the null detector II common lead to the low terminal of the TI, and wait until the thermal emf
caused by the heat from the operator's hands has subsided.
4.3.7 Set the null detectors operate–zero switch–to zero.
4.3.8 Set the voltage sources for minimum output, and disconnect the test configuration.

4.4 10 VOLT DC DECADE TEST

4.4.1 Connect equipment as shown in Figure 4.
2.4.3
DC VOLTAGE
CALIBRATOR
GUARD SENS OUT
2.5
VOLTAGE DIVIDER
INPUT
1.1 1.0 LO
HI LO
4.4.2 Set the TI voltage output control(s) to obtain a 10V dc output and the voltage divider control(s) to
0.999999X.
Voltage Divider
Setting
0.000999X
0.000699X
0.000499X
0.000199X
0.000099X
0*
OUTPUT
Figure 4. DC Volts Decade Test Configuration
NAVAIR 17–20AQ–70
Tolerance
Limits
2 V or less
"
"
"
"
3 V or less
TI
+
2.4.1
NULL DETECTOR
(BATTERY OPERATED)
IN
COM
GUARD
9