Required Equipment

The equipment in Table 2 is necessary for performance tests and calibration
adjustment.
Recommended Equipment
Fluke 5522A Calibrator
(or equivalent)

Performance Tests

To prevent possible electrical shock, fire, or personal injury,
do not perform the performance test procedures unless the
Product is fully assembled.
The performance tests verify the full operation of the Product and measure the
accuracy of each function against Product specifications. If the Product fails a
part of the test, calibration adjustment and/or repair is necessary. See
"Calibration Adjustment".

Test the Display

To verify that all segments of the display function:
1. With the Product off, push and hold .
2. Push  while you keep  pushed until all of the display segments are
shown. See Figure 1.
If segments of the display are missing, repair is necessary. See "Contact Fluke".
Table 2. Required Equipment
Function
DC Volts
Warning
Figure 1. All Segments of the Display