Cleaning the CF-300 and Accessories
WARNING
Do not disassemble or modify the CF-300 or AC adapter. Doing so may cause a fire or electric shock.
The CF-300 should not be operated if it is damaged, or if smoke or strange odors occur. Doing so
may result in a fire. In such situations, turn the power OFF immediately, disconnect the AC adapter
from the AC outlet, and contact the nearest Konica Minolta-authorized service facility.
CAUTION
When cleaning, disconnect the power plug. Connecting may cause electric shock.
Be careful not to pinch hands in parts of the instrument or accessories
that open and close.Doing so may result in injury.
Zero Calibration Box and White Calibration Plate
Wipe gently with a dry soft cloth. If dirt is difficult to remove, dampen a cloth with commercially available lens
cleaning liquid and wipe. Then wipe off the liquid with a cloth dampened with water, and leave it to dry.
c When cleaning, take care not to scratch the zero calibration box or white calibration plate.
Target Mask
Use a blower to remove dirt and dust from the target masks.
c Do not touch the white-coated surface of the target masks with fingers or wipe it with a cloth. If the target masks get
so dirty that dirt cannot be removed using a blower, contact the nearest Konica Minolta-authorized service facility.
Inside Integrating Sphere
1
. Set the specular component to SCI using the
software.
2
. Remove the target mask and protective cap,
then blow dust and dirt from the integrating
sphere opening with a blower.
{ Do not touch the white-coated inner surface of the
integrating sphere, wipe it with a cloth or place any
object against it. If the white-coated surface gets so
dirty that dirt cannot be removed using a blower,
contact the nearest Konica Minolta-authorized service
facility.
26