Cleaning the CM-3600A and Accessories

WARNING
Do not disassemble or modify the CM-3600A or AC adapter.
Doing so may cause a fire or electric shock.
The CM-3600A should not be operated if it is damaged, or if smoke or strange odors occur. Doing
so may result in a fire. In such situations, turn the power OFF immediately, disconnect the AC
adapter from the AC outlet, and contact the nearest Konica Minolta-authorized service facility.
CAUTION
When cleaning, disconnect the power plug. Connecting may cause electric shock.
Be careful around openings in the CM-3600A. Failure to do so
may result in fingers being trapped causing injury.

Zero Calibration Box and White Calibration Plate

Wipe gently with a dry soft cloth. If dirt is difficult to remove, dampen a cloth with commercially available lens
cleaning liquid and wipe. Then wipe off the liquid with a cloth dampened with water, and leave it to dry.
○ When cleaning, take care not to scratch the zero calibration box or white calibration plate.

Target Mask

Use a blower to remove dirt and dust from the target masks.
○ Do not touch the black-coated surface of the target masks with fingers or wipe it with a cloth. If the target
masks get so dirty that dirt cannot be removed using a blower, contact the nearest Konica Minolta-
authorized service facility.

Inside Integrating Sphere

1.
Make sure that there is nothing placed
against the illumination window.
2.
Block the receiving window so
that no dust or dirt enters.
3.
Open the sample holder and remove
dust and dirt using a blower.
○ Do not touch the white-coated inner surface
of the integrating sphere, wipe it with a cloth
or place any object against it. If the white-
coated surface gets so dirty that dirt cannot be
removed using a blower, contact the nearest
Konica Minolta-authorized service facility.

Receiving Window

1.
Set the measurement area to SAV using your
software.
Memo
• This will cause the lens to move closer to the
receiving window, making it easier to clean.
2.
Use a blower to remove dirt and dust from the
receiving window.
○ Do not put your fingers into the receiving
window or touch any optical lenses.
Illumination window
Receiving window
24
Receiving window