Revision A
Field Replaceable Units and Upgrades: Main Unit Components
5. Using a small, flat-blade screwdriver, remove the connector from the
NBP assembly to the DAS assembly.
3
6. If you need to replace the DAS assembly, attach the NBP assembly to
the new DAS assembly.
7. If you need to replace the NBP assembly, attach the new NBP
assembly to your unit's DAS assembly.
8. Reassemble the DAS and NBP assemblies. Make sure you do not
forget:
the four mounting screws,
the tube connection, and
the flex connector.
Dash 3000/4000 Patient Monitor
2000966-339
527A
DAS Assembly FRU
NBP Assembly FRU
526A
8-25