Revision A
Field Replaceable Units and Upgrades: Main Unit Components
6. Flip over the PCB and remove the PC Card socket from the edge
connector.
7. With the PC Card socket removed, place the HEX (M2) nuts back
into the PC Card socket (if they fell out earlier). Slide the RF LAN
shield around the PC Card socket. Use care not to bump the nuts out
of their position.
8. Be sure to have the RF shield in place so the holes line up on the
underside. The HEX nuts are captured in the PC Card socket and are
held into place by the shield.
Dash 3000/4000 Patient Monitor
2000966-339
553A
554A
555A
8-35