4-48
Maintenance: Checkout Procedures
6. Test the accuracy of these PPR settings:
Simulator Setting
70
100
160
7. Return the simulator to these conditions:
Set the SpO
% to 99.
2
Set the PPR to 100 beats/min.
8. Press the NOISE TEST button ON the simulator for 30 seconds.
9. Make sure the monitor still displays an SpO
100% (97 and 102%), or an interference detection message is
displayed.
10. Turn the simulator OFF.
11. Disconnect the simulator cable from the device under test.
Dash 3000/4000 Patient Monitor
2000966-339
Accuracy of PPR Settings
Displayed PPR Value
value between 97 and
2
68 – 72
97 – 103
156 – 164
Revision A