Field Replaceable Units and Upgrades: Main Unit Components
28. Reconnect the display flex to the DAS assembly. During assembly,
verify the RF LAN coaxes are not pinched between the front bezel
assembly and the rear housing.
29. Reattach the display assembly to the monitor's rear housing. It is
helpful to lay the monitor face down on a non-abrasive, static-free
surface. Make sure the Trim Knob control hangs over the edge of
the surface to prevent damage.
Top screws are M4 x 28L
Side screws are M4 x 14L
Bottom screws are M4 x 25L
30. Adhere RF LAN label to the rear of the monitor as shown.
Place RF LAN label here.
31. Reinstall the batteries and plug into AC power and verify that
Wireless LAN is enabled.
Dash 3000/4000 Patient Monitor
2000966-339
568A
584
8-41