Service Mode Menu Option Items
5-6
Troubleshooting: Service Menus
WARNING
Some of the service mode menu option items are to be
used only by qualified service technicians and others are
for general use. Because of this, unnecessary tampering
with service mode menu option items for experimentation
purposes is not recommended by GE and may cause a
malfunction of the monitor.
Following is a list of options in the main code service menu.
Menu Option
Review Errors
Calibrate
Battery Service
Patient-Monitor Type
WARNING
Changing the patient-monitor type will default the
admit function to Standard configuration. Different
alarms and parameters are activated for each
selection.
NOTE
The keypad/remote control is DIDCA programmed for specific
monitor types. The error message, "WARNING: REMOTE
MISMATCHED WITH MONITORING MODE" displays if the
monitor and keypad/remote control do not match.
Dash 3000/4000 Patient Monitor
2000966-339
Service Mode Menu Options
Description
This menu selection is for advanced troubleshooting
by GE engineers. Error log data can be transferred
over the network to a central station and then loaded
onto a diskette for review. (Review Errors is
discusses in greater detail later in this chapter.)
For checkout or calibration of the noninvasive blood
pressure, ECG analog output, BP analog output, CO
service, and SAM service menu functions of the
monitor.
This is a complete collection of battery data for
troubleshooting the batteries.
Select the type of monitor desired, i.e adult, neonatal
or operating room. Refer to Chapter 6,
"Configuration"
for detailed procedures.
2
Revision A