Opening the Display Assembly
8-16
Field Replaceable Units and Upgrades: Display Assembly
Complete the following steps for the monitor you are servicing.
Dash 3000 monitor
with display shield flex circuit
Separate the monitor bezel from the LCD
display.
1. Remove the four screws holding the
display shield to the LCD display and the
two screws holding the display shield to
the monitor bezel.
four larger screws
two smaller screws
2. Remove the display assembly from the
monitor bezel.
3. Pull the LCD display isolator out of the
way. Disconnect the LCD display cable.
LCD display cable
4. Disconnect the cable from the backlight
inverter PCB.
cable
Dash 3000/4000 Patient Monitor
2000966-339
Dash 4000 monitor with Sharp display
assembly
Separate the transition bezel from the front
bezel.
1. Remove the two screws from the display
flex cable. Pull up on the latch taps to
remove the flex cable from the display.
2 screws
2. Remove the six screws from the back of
the transition bezel and the four screws
500A
from the back of the metal display plate.
six screws
four screws
501A
3. Separate the transition bezel from the
front bezel.
4. Set aside the transition bezel and the
metal display plate.
502A
879A
880A
884A
Revision A