8-14
Field Replaceable Units and Upgrades: Display Assembly
c.
Carefully bend the display flex as shown.
d. Pick up and support the display assembly between the main unit
and a thick book.
e. Use two flathead M3X6LG screws from the kit to carefully secure
the flex processor cable connector to the main processor PCB.
f.
Carefully align and insert the the DAS cable pins into the DAS
ISO assembly until the snap latch clicks into place.
2 screws
Dash 3000/4000 Patient Monitor
2000966-339
2 screws
274A
875A
876A