8-36
Field Replaceable Units and Upgrades: Main Unit Components
9. Install the PC Card socket with the attached RF LAN shield into the
card edge connector on the PCB.
10. Flip over the PCB again. With the PC card socket in place, install the
four screws and tighten.
11. Install the seven screws from Step 1. Do NOT place a screw next to
J7.
12. Apply the supplied regulatory label below the Symbol PC card's
existing label. Refer to the RF LAN kit for the label.
13. Install Symbol PC card. Clip to hold card in place.
14. Install the expansion connector housing as done in Steps 2 and 3.
Dash 3000/4000 Patient Monitor
2000966-339
556A
557A
558A
Revision A