GE Dash 3000 Service Manual page 269

Patient monitor
Hide thumbs Also See for Dash 3000:
Table of Contents

Advertisement

GE Pulse Oximetry (SpO
PIN
SIGNAL NAME
1
NELLCOR_RCAL
2
IR/RED*
3
RED/IR*
4
NC
5
POX+
6
SHIELD
7
NELLCOR_RCAL_RETURN
8
MARQUETTE_PROBE*
9
POX–
10
NELLCOR_PROBE*
11
GND
Revision A
Assembly Drawings: Port Connections
) Cable Connector
2
The pulse oximetry function uses an 11-pin, female connector. The
pinout is as follows:
I/O
DESCRIPTION
O
Nellcor probe characteristics ID resistor
O
Anti-parallel LED drive
(low=RED, high=IR)
O
Anti-parallel LED drive
(low=IR, high=RED)
Not connected
I
Photo detector anode
Cable shield
I
Return for probe characteristics ID resistor
I
Marquette probe select
I
Photo detector cathode
I
Nellcor probe select
Ground reference for pins 8 and 10
Dash 3000/4000 Patient Monitor
2000966-339
DIAGRAM
587A
9-45