GE Dash 3000 Service Manual page 231

Patient monitor
Hide thumbs Also See for Dash 3000:
Table of Contents

Advertisement

Revision A
Assembly Drawings: Theory Of Operation
The DAS block diagram with GE SPO2 consists of the following three
sections.
Section
ECG/Respiration
Separately isolated section includes a 7.3728MHz 68HSC05
microcontroller, A/D conversion, signal processing hybrids and
DC-DC isolation converter.
Main DAS
Separately isolated section includes a 22.1184MHz 68332
microcontroller with FLASH and SRAM memory, A/D conversion,
signal processing hybrids and DC isolation converter.
Non-Isolated
Includes serial and parallel host interfaces and NBP pump, valves,
circuits
and over-pressure circuitry.
ECG
The ECG function detects heartbeats and arrhythmias, measures heart
rate (HR) and ST segment deviation, and generates a 12SL diagnostic
interpretation. Patient alarms with adjustable high and low limits for
HR and ST segment deviation are provided. Additional patient alarms
are provided for arrhythmias and PVCs. System alarms for individual
lead failure and all leads failure are provided.
The monitor accepts the green 3, 5, and 10-leadwire Multi-link ECG
connectors (compatible with Eagle 3000 monitor, Eagle 4000 monitor,
and Tram modules).
Respiration
The respiration function measures respiration rate (RR) and detects
apnea through the ECG leadwires using the impedance variation
technique. Patient alarms for RR (with adjustable high and low limits)
and apnea (with adjustable time limit) are provided. System alarms for
lead failure, cardiac artifact, and learning are provided.
GE Pulse Oximetry (SpO
The pulse oximetry function measures arterial oxygen saturation (SpO
and peripheral pulse rate (PPR). Patient alarms with adjustable high
and low limits for SpO
2
off patient, low-quality signal, and pulse search are provided.
The monitor accepts the blue color-coded pulse oximetry connector
(compatible with Eagle 3000 monitor, Eagle 4000 monitor, and the Tram
x50-series modules). The monitor with GE SPO2 supports Nellcor probes.
Dash 3000/4000 Patient Monitor
2000966-339
Description
)
2
and PPR are provided. System alarms for probe
)
2
9-7