Assembly Drawings: Theory Of Operation
Non-Invasive Blood Pressure
The NBP function measures systolic pressure, diastolic pressure, mean
pressure, and heart rate. Patient alarms with adjustable high and low
limits for systolic, diastolic, and mean pressures are provided. System
alarms for deflation failure, inflation failure, maximum pressure
exceeded, measurement time exceeded, pulse too weak, hardware
malfunction, and system pressure leak are provided.
The NBP function operates in manual, auto, and stat measurement
modes. The monitor has backup protections for magnitude and duration
of applied cuff pressure (with different settings in adult and neonatal
modes).
The monitor accepts the rectangular NBP connector (compatible with the
Eagle 3000 monitor and some versions of the Tram module).Invasive
Pressure
The invasive pressure function measures two blood pressures and
calculates systolic pressure, diastolic pressure, mean pressure, and
pulsatile pressure rate where applicable. Patient alarms with adjustable
high and low limits for systolic pressure, diastolic pressure, mean
pressure, and pulse rate are provided for each channel. System alarms
for sensor status (failure and disconnected), Smart BP event (artifact),
zeroing status (not zeroed, failure, and pressure sensed), and PA Wedge
status (wait, inflate, processing, complete, and no pulse) are provided.
The user can set an adjustable low-pass filter to 12 or 40 Hz. The 12 Hz
filter is implemented in software; the filter is disabled at the 40 Hz
setting.
The monitor accepts the red color-coded invasive pressure connectors
(compatible with the Eagle 3000 monitor, Eagle 4000 monitor, and Tram
modules).
Temperature
The temperature function measures two temperatures. Patient alarms
with adjustable high and low limits for temperature are provided.
System alarms for sensor and calibration failures are provided.
The monitor accepts the brown color-coded connector (compatible with
the Eagle 3000 monitor, Eagle 4000 monitor, and Tram modules). The
monitor supports both YSI Series 400 and 700 thermistor probes. The
probe type is determined by identification signals in the probe adapter
cable.
The temperature connector and measurement circuits are shared with
the cardiac output monitoring function; therefore you cannot use both
functions concurrently. A signal in the patient cable indicates the
appropriate function.
9-8
Dash 3000/4000 Patient Monitor
Revision A
2000966-339