8-24
Field Replaceable Units and Upgrades: Main Unit Components
option is installed, use a needle nose pliers to remove the antennae
by disconnecting both coax connectors from the PC card.
NOTE
If you need to replace the main assembly, the power supply
assembly, and/or the speaker, go to "Main and/or Power Supply
Assemblies, Speaker" later in this chapter.
3. Remove the four screws and remove the NBP assembly from the
cover of the DAS assembly.
4. Disconnect the tube from the inner solenoid going into the NBP
assembly.
four screws holding the NBP PCB to the DAS assembly
Dash 3000/4000 Patient Monitor
2000966-339
525A
Disconnect this end of the tube.
570A
Revision A