12SL and ACI-TIPI ECG Test
Revision A
Maintenance: Checkout Procedures
Perform this test if your monitor uses the 12SL ACI-TIPI ECG option.
1. Set up the patient simulator as follows:.
Heart rate – 80 bpm.
Heart rate amplitude – 1.0 mV.
12SL ECG patient cable (5-leads with V leads) properly attached.
2x gain for a MarqI, MarqII, or MarqIII simulator.
2. Select ECG from the monitor menu. Then, select 12 LEAD ECG
ANALYSIS.
3. Verify that the monitor is displaying 10 noise-free leads.
4. Select 12LD ECG NOW. Wait for the monitor to acquire and analyze
the data.
5. Select Transmit-Print.
6. Verify the 12SL ECG prints at the graph location assigned in the
monitor's Graph Setup >12SL Graph Location.
If there is no graph location is assigned, an error message
appears on the bottom of the monitor's display.
7. Verify the ECG is transmitted to the MUSE Cardiovascular
Information System. Verify the ECG prints out correctly as defined
by the MUSE system.
If no MUSE system is connected, an error message appears on
the bottom of the monitor's display.
8. Delete this test 12SL ECG from the MUSE system's edit list.
Dash 3000/4000 Patient Monitor
2000966-339
4-39