2-16
Equipment Overview: Technical Specifications
Cuff pressure range:
Adult:
Pediatric:
Neonatal:
Pressure accuracy:
Static:
Clinical:
Heart rate detection:
Total cycle time:
Automatic cycle times:
Auto zero:
Tubing length:
Adult:
Neonatal:
Automatic cuff deflation:
Cuff sizes:
Disposable:
Reusable:
Alarms:
Dash 3000/4000 Patient Monitor
2000966-339
294 ± 6 mmHg
250 ± 5 mmHg
147 ± 3 mmHg
±2% or ±3 mmHg, whichever is greater
±5 mmHg average error
8 mmHg standard deviation
30 to 200 beats per minute
20 to 40 seconds typical (dependent on heart rate and
motion artifact)
0 to 8 hours
Zero pressure reference prior to each cuff inflation
12 feet
8 feet
Cycle time exceeding 3 minutes (90 seconds neonatal),
power off, or cuff pressure exceeds 294 mmHg (±6
mmHg) adult, 147 mmHg (±3 mmHg) neonatal
Large adult, adult, small adult, pediatric, small pediatric,
and infant
Thigh, large adult, adult, child, and infant
User-selectable upper and lower limits for systolic,
diastolic, and mean pressures
Revision A