8-40
Field Replaceable Units and Upgrades: Main Unit Components
25. Reinstall DAS assembly. When installing the DAS assembly, it is
critical not to pinch the coax's or alarm light cable. Route alarm light
cable behind the DAS and off to the side of the upper left mounting
screws as shown in the picture below. Use four screws (M3 x 12L
with washers) to mount the DAS.
alarm light cable
26. Use needle nose pliers to snap the two coax connectors into the
Symbol PC card. Route coaxes per the picture. Note that the left coax
is routed above the RF LAN shield retaining clip.
27. Reinstall Display assembly. Use two (M3 x6L) screws to mount the
flex to the Processor/Power management PCB.
Dash 3000/4000 Patient Monitor
2000966-339
568A
567A
Revision A