5-26
Troubleshooting: Wireless LAN Troubleshooting
The most common cause of non-communication is an incorrect SSID. If
the communication status in the Wireless LAN ID is 01 (US 01 V4.63
00022), the RF LAN card is not communicating with the Access Point.
The SSID must match the Access Point's NET ID.
1. To change the SSID status:
a. Activate the Boot Code by holding down NBP GO/STOP and
ZERO ALL.
b. Press and release the Trim Knob control.
c.
Keep holding NBP GO/STOP and ZERO ALL until the Boot
Code information appears on the display.
d. Select SERVICE MENU > SET CONFIGURATION >
CONFIGURE WIRELESS LAN >CHANGE SSID.
NOTE
For all standard installations, the SSID should be GEMS
(default).
2. If the status is 02, the RF LAN card is communicating with the
Access Point and the network should be available.
a. Check the network connection to the Access Point.
b. Refer to the Wireless LAN (Symbol Access Point) Installation
and Service Manual for details on proper Access Point
installation.
Intermittent communication may be caused if the monitor is out of range
or there is RF interference.
If the monitor is out of range, add additional Access Points to extend
the coverage area.
If there is RF interference, the source of the interference must be
removed or shielded.
Dash 3000/4000 Patient Monitor
2000966-339
Revision A