8-34
Field Replaceable Units and Upgrades: Main Unit Components
4. Flip over the processor/power management PCB to expose the
backside of the PCB.
5. Remove the four (M2 x 12L) screws that hold on the PC card socket to
the Processor/Power Management PCB. Be careful not to drop the
four (M2) nuts captured in the PC card socket when removing these
screws. Pull the screws out of the PCB and set aside.
Dash 3000/4000 Patient Monitor
2000966-339
551A
Revision A