GE Dash 3000 Service Manual page 67

Patient monitor
Hide thumbs Also See for Dash 3000:
Table of Contents

Advertisement

Identifying the Battery Charge Capacity Using the Monitor
Revision A
Maintenance: Lithium-Ion Battery
One new, fully charged battery lasts about 2.5 hours. Two new, fully
charged batteries last about 5 hours.
A battery that has 100% Design Capacity and is charged to 100% of this
capacity lasts approximately 2.5 hours. However, as the Full Charge
Capacity declines, the approximate run time of a fully charged battery
will decrease.
Single Battery Run Time: Approximately 2.5 Hours
An icon fully outlined with solid lines and completely filled indicates a
new battery, fully charged to its Design Capacity.
Design Capacity and Full Charge Capacity
857A
Single Battery Run Time: Approximately 2.0 Hours
The dotted portion of the icon outline shows that the battery has lost 20%
of its Design Capacity. The solid-outline portion is filled in, showing that
the battery is charged to 100% Full Charge Capacity.
Full Charge Capacity
858A
Single Battery Run Time: Approximately 1.0 Hours
The dotted portion of the icon outline shows that the battery has lost 20%
of its Design Capacity. The solid-outline portion is half filled in, showing
that the battery is charged to only 50% of available capacity.
Full Charge Capacity
859A
50% Available Capacity
Dash 3000/4000 Patient Monitor
2000966-339