8-38
Field Replaceable Units and Upgrades: Main Unit Components
18. Re-install the battery door Assembly.
19. Install writer bracket in place. Connect the speaker cable to the
writer flex circuit. Connect the writer flex to J7 on the Processor/
Power Management PCB and secure it with the thumbscrew.
NOTE
When securing the thumb screw, first finger-tighten the thumb
screw. Then, use a flat-head screw driver to tighten the screw an
additional 1/4 turn.
20. Install writer (if equipped) or blank plate. (Writer option shown
here.)