GE Dash 3000 Service Manual page 190

Patient monitor
Hide thumbs Also See for Dash 3000:
Table of Contents

Advertisement

8-10
Field Replaceable Units and Upgrades: Display Assembly
4. Partially separate the display assembly from the main unit and look
inside the unit to determine how the display flex cable is "folded."
The location of the fold dictates how to position the display assembly
to remove it from the main unit.
CAUTION
CABLE DAMAGE — Do not kink, pinch, stretch, twist,
or tightly fold the display flex cable. If you do not handle
the flex cable gently, you will damage the cable.
Dash 4000 with the NEC display assembly (green-colored display shield)
Dash 3000 monitors
Dash 4000 with the Sharp display assembly (gold-colored display shield)
Dash 3000/4000 Patient Monitor
2000966-339
Flex cable folds up
from the bottom of the
display assembly.
869A
Flex cable folds over
from the right side of
the display assembly.
870A
Revision A