Copy Stored Monitor Defaults
Revision A
Configuration: Advanced User Procedures
WARNING
Loss of Defaults. Copying monitor defaults from the
server monitor erases all of the monitor defaults on the
client monitor.
To copy monitor defaults from the server monitor, the client monitor
must be in the Discharged state. If the client monitor is set to the OR
patient-monitor type, it must not be in Combo or Rover Combo Admit
mode.
NOTE
Make sure that the client monitor shares the same configuration as
the server monitor. Refer to
16.
1. At the client monitor, start from the Main Menu and select MORE
MENUS > MONITOR SETUP > SERVICE MODE.
2. Enter password using the Trim Knob control to select the day and
month from monitor screen with leading zeros. (e.g. July 4 = 0407)
3. Select COPY UNIT DEFAULTS.
4. Select SELECT UNIT TO COPY MONITOR DEFAULTS FROM.
5. From the list of units on the Unity network, choose the unit the
server monitor is in.
6. Select SELECT BED TO COPY MONITOR DEFAULTS FROM.
7. Scroll through the list of beds within the selected unit until the
desired server monitor is found. Select it, and answer YES to the
confirmation popup menu.
NOTE
After copying monitor defaults, the first set of defaults is
automatically activated. If another set of defaults is desired, the user
must manually select it from the Recall Defaults menu.
After copying monitor defaults from another bed (the server monitor),
verify that the defaults were transferred and arrhythmia levels are as
desired.
Dash 3000/4000 Patient Monitor
2000966-339
"Transfer Monitor Defaults" on page 6-
6-17