Masimo SET Pulse Oximetry (SpO
PIN
SIGNAL NAME
1
DETECTOR_ANODE
2
DETECTOR_CATHODE
3
NC
4
NC
5
IR_LED_CATHODE
6
OUTER_SHIELD
7
INNER_SHIELD
8
NC
9
NC
10
RED_LED_CATHODE
11
9-46
Assembly Drawings: Port Connections
) Cable Connector
2
The pulse oximetry function uses an 11-pin, female, "Nicolay-style"
pulse oximetry connector. Special keying only allows the use of Masimo
LNOP (low noise) probes. The pinout is as follows:
I/O
DESCRIPTION
I
Photodetector
I
Photodetector cathode
Not connected
Not connected
O
Anti-parallel LED drive
Inner cable shield - surrounds Detector_Anode and
Detector_Cathode signal pair. Terminated directly
to GND_ISO_MAIN via both wide copper traces
and the silver epoxy shield layer of the flex circuit.
Inner cable shield - surrounds Detector_Anode and
Detector_Cathode signal pair. Terminated directly
to GND_ISO_MAIN via both wide copper traces
and the silver epoxy shield layer of the 2006782-
001 flex circuit.
Not connected
Not connected
O
Anti-parallel LED drive
Outer cable shield - surrounds entire SpO2 cable.
Terminated directly to GND_ISO_MAIN via silver
epoxy shield layer of 2006782-001 flex circuit.
Dash 3000/4000 Patient Monitor
2000966-339
DIAGRAM
541A
Revision A