GE Dash 3000 Service Manual page 283

Patient monitor
Hide thumbs Also See for Dash 3000:
Table of Contents

Advertisement

Appendix A – Electromagnetic Compatibility (EMC): Electromagnetic Compatibility (EMC)
Guidance and Manufacturer's Declaration – Electromagnetic
Immunity
Immunity Test
EN 60601 Test Level
Conducted RF
EN 61000-4-6
150 KHz to 80 MHz
Radiated RF
EN 61000-4-3
80 MHz to 2.5 GHz
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by reflection from structures, objects,
and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radio,
AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the equipment is used exceeds the applicable RF compliance level above,
the equipment should be observed to verify normal operation. If abnormal performance is observed, additional measures
may be necessary, such as re-orienting or relocating the equipment.
b
Over the frequency range 150 KHz to 80 MHz, field strengths should be less than 3 V/m.
Revision A
The Dash 3000/4000 Patient Monitor is intended for use in the
electromagnetic environment specified below. It is the responsibility of
the customer or user to assure that the Dash 3000/4000 Patient Monitor
is used in such an environment.
Compliance Level
3 Vrms
3 V rms
3 V/m
3 V/m
Dash 3000/4000 Patient Monitor
Electromagnetic Environment – Guidance
Portable and mobile RF communications equipment should not
be used closer to any part of the equipment, including cables,
than the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance
P
d = 1.2
P
d = 1.2
80 MHz to 800 MHz
P
d = 2.3
800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer, and d is
the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey
compliance level in each frequency range
Interference may occur in the vicinity of equipment marked with
the following symbol:
2000966-339
a
, should be less than the
b
.
A-5