4-46
Maintenance: Checkout Procedures
5. Observe a reading of 200/200 (200) ± 4 mmHg (or 240 ± 4 mmHg if
using the Marq II or MarqIII) in the BP parameter window on the
monitor display.
6. Change the patient simulator BP output to WAVE or to a 120/80
wave form (simulated BP waveform).
7. Observe a distortion-free BP waveform and a reading of
approximately 120/80 (93) in the BP parameter window on the
monitor display.
8. Disconnect the BP simulator cable from the BP connector of the
monitor.
9. Complete steps for each enabled BP connector.
10. This completes the BP tests.
Dash 3000/4000 Patient Monitor
2000966-339
Revision A