5-8
Troubleshooting: Service Menus
Service Mode Menu Options (Continued)
Menu Option
WARNING
Duplication of an internet (IP) address on a network
causes lost data. If you change the factory assigned
internet address, you must first record all other
internet addresses used on your network to avoid
duplication.
NOTE
An incorrect internet address may also prevent the monitor
from viewing other monitors on the network even though the
unit names match. Whether or not this can occur depends on
the network topology at the installed site.
MONITOR SETTINGS
COPY UNIT DEFAULTS
GRAPH TEST PATTERN
TIME AND DATE
WARNING
Loss of patient history. This menu should rarely be
used because patient histories will be lost.
Dash 3000/4000 Patient Monitor
2000966-339
Description
STORE DEFAULTS FOR NETWORK TRANSFER
This menu selection saves the monitor defaults for
transfer to another monitor with the same software
version, software package, patient monitor type,
and country code. It stores all monitor defaults,
custom default names, and 12 SL location and site.
SET LOCATION ID
Enter the location identification for the MUSE
system.
SET SITE NUMBER
A site number may be necessary if your institution
has multiple buildings. A MUSE system can be
used as a site number.
This menu option is used to copy monitor defaults
from one monitor to another monitor. Refer to
Chapter 6,
"Configuration"
for detailed procedures.
This menu selection allows you to run a graph test
pattern. The choices are Start and Stop.
This menu selection allows changes to the time and
date and may affect the time and date for the entire
monitoring network. (Refer to Chapter 6,
"Configuration"
for detailed procedures.)