TCP/IP Parameter Configuration

To operate properly in a network environment, the LAN interface must be
configured with valid TCP/IP network parameters. These parameters are:
• IP address
• Subnet Mask
• Default Gateway
The TCP/IP parameters can be configured by the following methods:
• by automatically requesting the parameters from a network-based DHCP
• by manually setting the parameters using Telnet
• by manually setting the parameters using the Instant Pilot (G4208A)
The LAN interface differentiates between several initialization modes. The
initialization mode (short form 'init mode') defines how to determine the
active TCP/IP parameters after power-on. The parameters may be derived
from non-volatile memory or initialized with known default values. The
initialization mode is selected by the configuration switch, see
page 271.
1290 Infinity Quaternary Pump User Manual
Server (using the so-called Dynamic Host Configuration Protocol). This
mode requires a LAN-onboard Module or a G1369C LAN Interface card, see
"Setup (DHCP)"
on page 274
LAN Configuration

TCP/IP Parameter Configuration