Remove the Pump Head Assembly
6
Counter the lock screw of the heat exchanger capillary
while loosening the outlet valve. Keep the outlet valve
installed to the pump head assembly.
174
7
Open the four screws holding the pump heads.
N O T E
Open all screws step by step, not screw by screw.