Replacing The Multi Purpose Valve - Agilent Technologies 1290 Infinity User Manual

Quaternary pump
Hide thumbs Also See for 1290 Infinity:
Table of Contents

Advertisement

Replacing the Multi Purpose Valve

Tools required
p/n
5023-0240
Parts required
p/n
0100-1259
01080-83202
5067-4174
Preparations
Remove all capillary connections from the Multi Purpose Valve.
1
Remove the clamp with the inline filter (if installed).
1290 Infinity Quaternary Pump User Manual
Description
Hex driver, ¼", slitted
Description
Blank nut (plastic)
Blank nut (stainless steel)
Multi Purpose Valve Head
2
Unscrew the black union nut and remove the head of the
purge valve by pulling it to the front.
Maintenance

Replacing the Multi Purpose Valve

9
193