12
Hardware Information
Firmware Description
Some modules are limited in downgrading due to their main board version or their initial
N O T E
firmware revision. For example, a G1315C DAD SL cannot be downgraded below firmware
revision B.01.02 or to a A.xx.xx.
Some modules can be re-branded (e.g. G1314C to G1314B) to allow operation in specific
control software environments. In this case, the feature set of the target type is used and
the feature set of the original one is lost. After re-branding (e.g. from G1314B to G1314C),
the original feature set is available again.
All this specific information is described in the documentation provided with the firmware
update tools.
The firmware update tools, firmware and documentation are available from
the Agilent web.