3
Installing the Module
Installing the Pump
3 Connect the power cable to the power connector at the back of the module.
4 Connect the required interface cables to the rear of the pump.
In an Agilent 1290 Infinity System, the individual modules are connected by CAN cables.
N O T E
An Agilent 1200 Series Instant Pilot can be connected to the CAN bus of any module.
Connection to an Agilent data system is established through the built-in LAN port of the
detector. The LAN port of the detector must be used as the detector generates the highest
data rate of all modules. For more information about connecting the Instant Pilot or Agilent
Data System, please refer to the respective user manual. For setting up the LAN access,
see
5 Turn on the power by pushing the button at the lower left hand side of the
When the line power button stands out and the green light is off, the module is turned off.
N O T E
The module was shipped with default configuration settings. For changing these settings,
N O T E
refer to section Setting the 8-bit configuration switch.
42
Figure 14
Rear view of the quaternary pump
"LAN Configuration"
on page 267.
module.
The power button stays pressed in and the status LED should be green.
1290 Infinity Quaternary Pump User Manual