Agilent Technologies 1290 Infinity User Manual page 152

Quaternary pump
Hide thumbs Also See for 1290 Infinity:
Table of Contents

Advertisement

9
Maintenance
Replacing the Inlet Weaver
3
Pull the inlet weaver out of the inlet valve.
5
Fix the fitting of the new inlet weaver to the MCGV.
152
4
Insert the new inlet weaver to the inlet valve. Fix the
weaver with the plastic screw.
6
Reconnect tubings between MCGV and degasser.