Agilent Technologies 1290 Infinity User Manual page 187

Quaternary pump
Hide thumbs Also See for 1290 Infinity:
Table of Contents

Advertisement

9
Remove the seal holders from the pump chambers.
11
Place the seal holder onto the spring housing.
N O T E
Mind the correct orientation: The backup ring must
face upwards and the seal holder must sit correcly.
1290 Infinity Quaternary Pump User Manual
Pump Head Maintenance (Tool Free)
10
Lubricate the piston with isopropanol and place it into
the spring housing.
12
Place the assembled spring housings on top of the pump
chambers.
N O T E
Both spring housings are identical, there is no risk
when mixing them, but make sure that the seal holder
is oriented correctly.
9
Maintenance
187