Introduction
System Overview
The solvent cabinet (1) is designed to store a maximum volume of 6 L solvent.
The maximum volume for an individual bottle stored in the solvent cabinet
should not exceed 2.5 L. For details, see the usage guideline for the Agilent
1200 Infinity Series Solvent Cabinets (a printed copy of the guideline has been
shipped with the solvent cabinet, electronic copies are available on the
Internet).
The leak pan (2) (individually designed in each module) guides solvents to the
front of the module. The concept covers also leakages on internal parts
(e.g. the detector's flow cell). The leak sensor in the leak pan stops the running
system as soon as the leak detection level is reached.
The leak pan's outlet port (3, A) guides excessive overfill from one module to
the next, as the solvent flows into the next module's leak funnel (3, B) and the
connected corrugated waste tube (3, C). The corrugated waste tube guides the
solvent to the next lower positioned module's leak tray and sensor.
The waste tube of the sampler's needle wash port (4) guides solvents to waste.
The condense drain outlet of the autosampler cooler (5) guides condensate to
waste.
The waste tube of the purge valve (6) guides solvents to waste.
The waste tube connected to the leak pan outlet on each of the bottom
instruments (7) guides the solvent to a suitable waste container.
20
1290 Infinity Quaternary Pump User Manual