Agilent Technologies 1290 Infinity User Manual page 180

Quaternary pump
Hide thumbs Also See for 1290 Infinity:
Table of Contents

Advertisement

9
Maintenance
Pump Head Maintenance (Tool Free)
13
Pull out the Seal Handling Device with the piston seal in
a straight movement with only gentle force.
15
Screw the pin of the seal handling device into the wash
seal.
N O T E
The seal holder has two different sides. The black
backup ring is supporting the piston seal and must
not be removed. The side with the backup ring has
a bigger diameter and a sharp edge to hold the
piston seal. The other side has no sharp edge and
holds the smaller wash seal.
180
14
Repeat for the other pump chamber.
16
Pull out the Seal Handling Device with the wash seal in a
straight movement with only gentle force.
1290 Infinity Quaternary Pump User Manual