1
Remove the capillaries from the Multi Purpose Valve to
the inline filter.
3
Use two 5/16 wrenches for opening the inline filter.
1290 Infinity Quaternary Pump User Manual
Replacing Parts of the Inline Filter
2
Remove the inline filter from the clamp attached to the
Multi Purpose Valve.
4
Replace the filter frit and reassemble the inline filter.