Agilent Technologies 1290 Infinity User Manual page 275

Quaternary pump
Hide thumbs Also See for 1290 Infinity:
Table of Contents

Advertisement

2 Set the Configuration Switch to DHCP either on the G1369C LAN Interface
Table 17
SW 4
ON
Table 18
SW 6
ON
3 Turn on the module that hosts the LAN interface.
4 Configure your Control Software (e.g. OpenLAB CDS ChemStation Edition,
1290 Infinity Quaternary Pump User Manual
Card or the main board of above mentioned modules.
G1369C LAN Interface Card (configuration switch on the card)
SW 5
SW 6
OFF
OFF
LC Modules with 8-bit configuration switch (B-firmware) (configuration switch
at rear of the instrument)
SW 7
SW 8
OFF
OFF
Lab Advisor, Firmware Update Tool) and use MAC address as host name,
e.g. 0030d3177321.
The LC system should become visible in the control software (see Note in
section
"General Information (DHCP)"
Dynamic Host Configuration Protocol (DHCP)
SW 7
SW 8
Initialization Mode
OFF
OFF
DHCP
Initialization Mode
DHCP
on page 273).
13
LAN Configuration
275

Advertisement

Table of Contents
loading

Table of Contents