Replace The Heat Exchanger - Agilent Technologies 1290 Infinity User Manual

Quaternary pump
Hide thumbs Also See for 1290 Infinity:
Table of Contents

Advertisement

9
Maintenance
Pump Head Maintenance (Tool Free)

Replace the Heat Exchanger

Tools required
p/n
5023-2501
5067-5688
G4220-20013
G4220-20015
G4220-20041
Parts required
p/n
G4220-81013
Preparations
• Remove the pump head assembly from the pump
• Remove the secondary pump head from the link plate
Loss of small spacer fitting
C AU T I O N
Inside the secondary pump head is a small spacer fitting, which can be dropped easily
when removing the heat exchanger.
1
Remove the 19 mm screw at the front of the secondary
pump head.
182
Description
Wrench, 19 mm
Screwdriver Torx-T10
Torque wrench 1 – 25 Nm with 14 mm wrench
4 mm hex bit
Adapter ¼ in square to hex
Bit Torx 10x25 mm
Description
Heat Exchanger Channel A (secondary pump head only)
The heat exchanger does not need to be removed for pump head maintenance.
2
Remove the front plate.
1290 Infinity Quaternary Pump User Manual