Agilent Technologies 1290 Infinity User Manual page 188

Quaternary pump
Hide thumbs Also See for 1290 Infinity:
Table of Contents

Advertisement

9
Maintenance
Pump Head Maintenance (Tool Free)
13
Place the screws into the pump heads and loosely
tighten them in a crosswise manner.
N O T E
The spring housing will tilt slightly when the first
screw is hand tightened. Stop at this point and
continue to tighten the three other screws in a
crosswise manner.
15
Tighten the pump head screws with a torque wrench set
to approx. 5 Nm in a crosswise manner.
N O T E
When the wrench clicks, the set torque is reached. Do
not overtighten the screws.
188
14
Mind the correct orientation of the link plate and click it
into place.
N O T E
The Pump Head Holder has a marker to illustrate the
correct placement of the link plate. The link plate
holds an identification tag; this has to be placed onto
the correct position to be readable by the pump.
16
Mount the pump head to the module. Do not fix the
screws at this stage!
1290 Infinity Quaternary Pump User Manual

Advertisement

Table of Contents
loading

Table of Contents