Error Information
Pump Error Messages
Drive phases differ too much in electric resistance
Error ID: 29216
Pump drive calibration has failed due to a strong difference electric
resistances for different motor phases.
Probable cause
1
2
3
Degasser's pressure limit violation
Error ID: 29220
Pressure too far above the limit.
Probable cause
1
2
3
130
Pump drive cable defect.
Pump drive defect.
Defective main board.
Leak in degasser chamber or degasser
tubing.
Defect vacuum pump.
Degasser chamber empty or connected to
air.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Block unused degasser channels.