Table 2
Type
Composition accuracy
Integrated degassing unit
Control
Local control
Communications
Safety and maintenance
GLP features
Housing
1290 Infinity Quaternary Pump User Manual
Performance specifications
Specification
±0.40 % absolute (1 – 99 % B,
0.5 – 2.0 mL/min with water/caffeine
tracer, 400 bar)
Number of channels: 4
Internal volume per channel: 1.5 mL
Agilent ChemStation for LC (C.01.04 or
above)
OpenLAB (A.04.04)
Masshunter (B.05.01 or above)
Agilent Instant Pilot (G4208A) (B.02.08 or
above)
Controller-area network (CAN), RS232C,
APG remote: ready, start, stop and
shutdown signals, LAN
Extensive diagnostics, error detection and
display through Agilent LabAdvisor, leak
detection, safe leak handling, leak output
signal for shutdown of the pumping
system. Low voltage in major maintenance
areas.
Early maintenance feedback (EMF) for
continuous tracking of instrument usage in
terms of seal wear and volume of pumped
mobile phase with pre-defined and user
settable limits and feedback messages.
Electronic records of maintenance and
errors.
All materials recyclable.
Site Requirements and Specifications
Performance Specifications
Comments