2.3.3 Step 3 - Adjust Level,
Parameters
calculate flow. In this flow conversion mode, the flow meter uses
internal look-up tables for many common primary measuring
devices.
An Equation is used when you have a non-standard primary
device, or want to use different values from those programmed
into the look-up tables of the flow meter. Equation uses the
standard flow equation:
P1
Q = k1H
+ k2H
Where Q equals flow rate; k1 and k2 are constants; H is level (or
head), and P1 and P2 are the powers to which the two H terms
are raised. (Your equation may not have the second term, in
which case you would enter 0 for the second constant, k2.) Most
common primary devices are supported in the flow meter's
software, so generally you will not need this option. But it is
available for those needing to enter their own values, or for those
who have a nonstandard primary device for which an equation
can correlate level and flow.
MANNING Flow Conversion uses the Manning formula to cal-
culate flow in open or closed (non pressurized) gravity-flow situa-
tions based on slope, diameter, and roughness of the pipe. The
Manning formula is named for its developer, Robert Manning, a
19th-century Irish civil engineer. There is no primary measuring
device as such. Instead the pipe, with considerations for its slope
and internal roughness, serves as the primary device. The 4250
Flow Meter can calculate flow in round pipes, rectangular,
U-shaped, or trapezoidal channels based on this formula.
Data Point Flow Conversion (DATA POINTS) calculates flow
based on a set of user-entered data points for a flow stream. Data
consist of correlated level and flow measurements for the stream.
Like the Equation method of flow conversion, this flow con-
version is most commonly used where the primary measuring
device is nonstandard, but where tables of level and flow rate
data are available from the device manufacturer. The 4250 Flow
Meter has space for four sets of data with as many as fifty points
per set. The flow meter then calculates flow from these data
tables using a three-point interpolation.
Adjust Level, Parameters calibrates the measuring sensors
that provide the flow meter with level and other information. In
this step you set the level that the flow meter measures. First
you measure, as accurately as possible, the level in the flow
stream. Then you enter this value with the numeric keys.
Note
Accuracy is important. The level measurement and channel
dimensions that you enter into the flow meter provide the basis
for all subsequent flow calculations.
The flow meter also has an input port for measurements other
than level. This is the Parameter Port. Here you can measure
such variables in the flow stream as temperature, pH (the acidity
Section 2 Programming
P2
4250 Flow Meter
2-7