Operating Parameters & Sensor Types; Typical Operating Parameters - Teledyne PS200 Series Operation & Maintenance Manual

Portable gas monitor
Table of Contents

Advertisement

PS200
PORTABLE GAS MONITOR
OPERATING PARAMETERS & SENSOR TYPES
A. OPERATING PARAMETERS & SENSOR TYPES
A.1. Typical Operating Parameters
Gas
LEL
Oxygen (O
)
2
Carbon Monoxide (CO)
Hydrogen Sulfide (H
S)
2
Dimensions (H x W x D):
Weight with Pump:
Weight with Pump:
Operating Temperature
Limits:
Storage Temperature Limits:
Humidity:
Nominal Flow Rate:
Typical Flow Fail Rate:
~30 seconds (~90 seconds for monitors with IR flammable sensor options)
Visual:
Audible:
Green/Red LCD backlit display
Battery:
Runtime:
Runtime Using Lo Power LEL
Sensor:
Charging Time:
High-impact, rubberized polycarbonate case. Can withstand physical impact testing to EN 60079 section 1-5
IP67
Timed:
Session:
Calibration:
Bump:
*Based on 1 gas event every 1 hour during a typical 8-hour shift, with a datalogging interval of 1 minute.
64171, Rev. 11/December 2019
Range
0 to 100%
1%
0 to 25%
0.1%
0 to 1000 ppm
1 ppm
0 to 9.9 ppm
0.1 ppm
10 to 100 ppm
1 ppm
Physical Properties
4.8" x 2.3" x 1.3" (121 mm x 59 mm x 32 mm)
8 oz. (230 g)
7.6 oz (215 g)
Environment
-4ºF to +122ºF (-20ºC to +50ºC)
-40ºF to +149ºF (-40ºC to +65ºC)
0 to 95% R.H non-condensing
0.132 to 0.185 gals / min (0.5 to 0.7 liters / min) Max. 97 ft. (30 m) sample line.
0.026 to 0.053 gals / min (0.1 to 0.2 liters / min)
Warm-up/Stabilization Time
Alarms
Highly visible, flashing LED
Sounder >90 dB
Display
Power Source
Rechargeable Lithium-Ion
Up to 14 hours (8 hours pumped)
Up to 80 hours (20 hours pumped)
Up to 4 hours
Construction
IP Rating
Datalogging
Minimum of 6 months datalogging*
Minimum of 360 logs
Minimum of 8 logs
Minimum of 360 logs
Resolution
10 secs. (IEC 60079-26-1)
12 secs. (BSEN 50104)
<20 secs.
<20 secs.
Response Time
A-1