Warm-up Time

T3LCR series is ready to be used as soon as the power-up sequence
has completed. However, to achieve the specification accuracy, first
warm up the instrument for 30 minutes.

Connect to the Device Under Test (DUT)

The T3LCR series uses a four terminal measurement configuration
that provides easy, accurate and stable measurements and avoids
mutual inductance, interference from measurement signals, noise
and other factors inherent with other types of connections.
Connect to DUT
WARNING
6
L Force
L Sense
-
Identifies conditions or practices that could result
in injury or loss of life. Do not apply DC voltage or
current to the UNKNOWN terminals. Applying a
DC voltage or current may lead to device failure.
Connect the measurement sample (DUT) to the
test port (or the test fixture, cables, etc. after the
DUT has been completely discharged.
H Sense
- DUT +
+
H Force